1. Soak the puncture rack in low foam with neutral pH and enzyme detergent for at least five minutes;
2. When soaking, use an instrument brush to brush off residual pollutants on the surface and components. If visible pollutants are difficult to remove, repeat the soaking procedure;
3. Remove the puncture stand from the cleaning solution and use it to wipe off all residues with a cloth;
4. Steam (high-pressure sterilizer) can be used to disinfect or sterilize the puncture frame at a temperature of 132-138 ℃ for 15 to 30 minutes;
1. A device used in conjunction with intracavitary or extracorporeal ultrasound probes for guiding and fixing puncture instruments such as puncture needles in ultrasound diagnosis and treatment operations;
2. Open the clip, align the positioning marks of the puncture frame and probe, and press down to place it on the probe. Lock the upper and lower clips (or naturally fasten with elastic force);
3. After checking that the puncture frame is securely installed, the puncture needle can be inserted into the guiding structure for puncture operation. Please refer to the manual or installation diagram for use